*RST Condition: FIFO is empty.
Usage
SENSe:DATA:FIFO?
Return all FIFO readings until
measurements complete and FIFO empty.
Command
Sequence
Set up scan lists and trigger.
SENSe:DATA:FIFO:ALL?
Now execute read statement.
Read statement does not complete until
triggered measurements are complete and
FIFO is empty.
[SENSe:]DATA:FIFO:COUNt?
[SENSe:]DATA:FIFO:COUNt?
returns the number of readings currently in the
FIFO buffer.
Comments
•
SENSe:DATA:FIFO:COUNt?
is used to determine the number of readings to
acquire with the
SENSe:DATA:FIFO:PART?
command.
•
Returned Value: Numeric
0
through
65,024
. The C-SCPI type returned is
int32.
•
Related Commands:
[SENSe:]DATA:FIFO:PART?
•
*RST Condition: FIFO empty.
Usage
SENSe:DATA:FIFO:COUNt?
Check the number of readings in the FIFO
buffer.
[SENSe:]DATA:FIFO:COUNt:HALF?
[SENSe:]DATA:FIFO:COUNt:HALF?
returns a
1
if the FIFO is at least half full
(contains at least 32,768 readings), or
0
if FIFO is less than half-full.
Comments
•
SENSe:DATA:FIFO:COUNt:HALF?
is used as a fast method to poll the FIFO for
the half-full condition.
•
Returned Value: Numeric
1
or
0
. The C-SCPI type returned is int16.
•
Related Commands:
[SENSe:]DATA:FIFO:HALF?
•
*RST Condition: FIFO empty.
Command
Sequence
SENSe:DATA:FIFO:COUNt:HALF?
Poll FIFO for half-full status.
SENSe:DATA:FIFO:HALF?
Returns 32768 readings.
[SENSe:]
Chapter 5
HP E1313/E1413 Command Reference 231