SCPI Command Quick Reference
Command
Description
:TARE (@<ch_list>)
Calibrate out system field wiring offsets.
:RESet
Resets cal constants from CALibration:TARE back to zero for all
channels.
:TARE?
Returns state of CALibration:TARE operation (returns error codes or 0
for OK).
:VALue
:RESistance <ref_ohms>
Send to instrument the value of just measured reference resistor.
:VOLTage <ref_volts>
Send to instrument the value of just measured voltage reference.
:ZERO?
Correct A/D for short term offset drift (returns error codes or 0 for OK).
DIAGnostic
:CALibration
:TARE
[:OTDetect]
[:MODE]
Sets mode to control OTD current during tare calibration.
[:MODE]?
Returns current setting of OTD control during tare calibration.
:CHECksum?
Perform checksum on Flash RAM and return a ’1’ for OK, a ’0’ for
corrupted or deleted memory contents.
:COMMand
:SCPWRITE <reg_addr>,<reg_data>
Writes values to SCP registers.
:CUSTom
:LINear <table_ad_range>,<table_block>,(@<ch_list>)
Loads linear custom EU table.
:PIECewise <table_ad_range>,<table_block>,(@<ch_list>)
Loads piecewise custom EU table.
:REFerence:TEMPerature
Puts the contents of the Reference Temperature Register into the FIFO.
:FLOor[:CONFigure] <range>,(@<ch_list>)
Sets lowest range that autorange can select.
:FLOor:DUMP
Sends the autorange floor setting for all 64 channels to the FIFO.
:INTerrupt[:LINe] <intr_line>
Sets the VXIbus interrupt line the module will use.
:INTerrupt[:LINe]?
Returns the VXIbus interrupt line the module is using.
:OTDetect[:STATe] ON | OFF, (@<ch_list>)
Controls “ Open Transducer Detect” on SCPs contained in <ch_list>.
:OTDetect[:STATe]? (@<channel>)
Returns current state of OTD on SCP containing <channel>.
:QUERy
:SCPREAD <reg_addr>
Returns value from an SCP register.
:VERSion?
Returns manufacturer, model, serial#, flash revision #, and date
e.g. HEWLETT-PACKARD,E1413C,US34000478,A.04.00,
Wed Jul 08 11:06:22 MDT 1994
FETCh?
Return readings stored in VME Memory (format set by FORMat:DATA
cmd).
FORMat
[:DATA] <format>[, <size>]
Set format for response data from [SENSe:]DATA?.
ASCii[, 7]
Seven bit ASCII format (not as fast as 32-bit because of conversion).
PACKed[, 64]
Same as REAL, 64 except NaN, +INF, and -INF format compatible
with HP BASIC.
REAL[, 32]
IEEE 32-bit floating point (requires no conversion so is fastest).
REAL, 64
IEEE 64-bit floating point (not as fast as 32-bit because of conversion).
[:DATA]?
Returns format: REAL, +32 | REAL, +64 | PACK, +64 | ASC, +7.
INITiate
:CONTinuous ON | OFF
When ON, module returns to Waiting for Trigger state after each scan.
When OFF, module goes to Trigger Idle state after each scan.
[:IMMediate]
Put module in Wait-for-Trigger state (ready to make one scan).
INPut
:FILTer
Control filter Signal Conditioning Plug-ons.
[:LPASs]
:FREQuency <cutoff_freq<D>,(@<ch_list>)
Sets the cutoff frequency for active filter SCPs.
:FREQuency? (@<channel>)
Returns the cutoff frequency for the channel specified.
[:STATe] ON | OFF, (@<channel>)
Turn filtering OFF (pass through) or ON (filter).
[:STATe]? (@<channel>)
Return state of SCP filters.
:GAIN <chan_gain>,(@<ch_list>)
Set gain for amplifier-per-channel SCP.
:GAIN? (@<channel>)
Returns the channel’s gain setting.
Command Quick Reference
Chapter 5
HP E1313/E1413 Command Reference 285
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com