Model 2461 Interactive SourceMeter® Instrument Reference Manual
Section 8: TSP command reference
2461-901-01 A/November 2015
8-331
tspnet.termination()
This function sets the device line termination sequence.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Function
No
Usage
type
= tspnet.termination(
connectionID
)
type
= tspnet.termination(
connectionID
,
termSequence
)
type
The termination type:
•
tspnet.TERM_LF
•
tspnet.TERM_CR
•
tspnet.TERM_CRLF
•
tspnet.TERM_LFCR
connectionID
The connection ID returned from
tspnet.connect()
termSequence
The termination sequence:
•
tspnet.TERM_LF
•
tspnet.TERM_CR
•
tspnet.TERM_CRLF
•
tspnet.TERM_LFCR
Details
This function sets and gets the termination character sequence that is used to indicate the end of a
line for a TSP-Net connection.
Using the
termSequence
parameter sets the termination sequence. The present termination
sequence is always returned.
For the
termSequence
parameter, use the same values listed in the table above for type. There are
four possible combinations, all of which are made up of line feeds (LF or 0x10) and carriage returns
(CR or 0x13). For TSP-enabled devices, the default is
tspnet.TERM_LF
. For devices that are not
TSP-enabled, the default is
tspnet.TERM_CRLF
.
Example
deviceID = tspnet.connect("192.0.2.1")
if deviceID then
tspnet.termination(deviceID, tspnet.TERM_LF)
end
Sets termination type for IP address
192.0.2.1 to
TERM_LF
.
Also see
(on page 8-325)
(on page 8-326)