Section 8: TSP command reference
Model 2461 Interactive SourceMeter® Instrument Reference Manual
8-216
2461-901-01 A/November 2015
smu.source.sweeplog()
This function sets up a logarithmic sweep for a set number of measurement points.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Function
Yes
Usage
smu.source.sweeplog(
configListName
,
start
,
stop
,
points
)
smu.source.sweeplog(
configListName
,
start
,
stop
,
points
,
sDelay
)
smu.source.sweeplog(
configListName
,
start
,
stop
,
points
,
sDelay
,
count
)
smu.source.sweeplog(
configListName
,
start
,
stop
,
points
,
sDelay
,
count
,
rangeType
)
smu.source.sweeplog(
configListName
,
start
,
stop
,
points
,
sDelay
,
count
,
rangeType
,
failAbort
)
smu.source.sweeplog(
configListName
,
start
,
stop
,
points
,
sDelay
,
count
,
rangeType
,
failAbort
,
dual
)
smu.source.sweeplog(
configListName
,
start
,
stop
,
points
,
sDelay
,
count
,
rangeType
,
failAbort
,
dual, bufferName
)
smu.source.sweeplog(
configListName
,
start
,
stop
,
points
,
sDelay
,
count
,
rangeType
,
failAbort
,
dual, bufferName
,
asymptote
)
configListName
A string that contains the name of the configuration list that the instrument will
create for this sweep
start
The voltage or current source level at which the sweep starts:
•
Current: 1 µA to 7.35 A
•
Voltage 200 mV to 105 V
stop
The voltage or current at which the sweep stops:
•
Current: 1 µA to 7.35 A
•
Voltage 200 mV to 105 V
points
The number of source-measure points between the start and stop values of the
sweep (2 to 1e6); to calculate the number of source-measure points in a sweep, use
the following formula:
Points = [(Stop - Start) / Step] + 1
sDelay
The delay between measurement points; default is
smu.DELAY_AUTO
, which
enables autodelay, or a specific delay value from 50
µ
s to 10 ks, or 0 for no delay
count
The number of times to run the sweep; default is
1
:
•
Infinite loop:
smu.INFINITE
•
Finite loop:
1
to
268,435,455
rangeType
The source range that is used for the sweep:
•
Most sensitive source range for each source level in the sweep:
smu.RANGE_AUTO
•
Best fixed range:
smu.RANGE_BEST
(default)
•
Present source range for the entire sweep:
smu.RANGE_FIXED
failAbort
Complete the sweep if the source limit is exceeded:
smu.OFF
Abort the sweep if the source limit is exceeded:
smu.ON
(default)
dual
Determines if the sweep runs from start to stop and then from stop to start:
•
Sweep from start to stop only:
smu.OFF
(default)
•
Sweep from start to stop, then stop to start:
smu.ON
bufferName
Indicates the reading buffer to use; the default buffers (
defbuffer1
or
defbuffer2
) or the name of a user-defined buffer; if no buffer is specified,
defbuffer1
is used
asymptote
Default is 0; see
Details