Appendix D: Convert Model 2400 to Model 2461 code
Model 2461 Interactive SourceMeter® Instrument Reference Manual
D-14
2461-901-01 A/November 2015
Model 2400 command
[:SENSe[1]]:AVERage:TCONtrol
[:SENSe[1]]:AVERage:TCONtrol?
Model 2461 command
[:SENSe[1]]:<function>:AVERage:TCONtrol
(on page 6-56), where
<function>
is
VOLTage[:DC]
,
CURRent[:DC]
, or
RESistance
.
Notes
This is now set for each measurement function.
Model 2400 command
[:SENSe[1]]:CURRent[:DC]:NPLCycles
[:SENSe[1]]:CURRent[:DC]:NPLCycles?
[:SENSe[1]]:RESistance:NPLCycles
[:SENSe[1]]:RESistance:NPLCycles?
[:SENSe[1]]:VOLTage[:DC]:NPLCycles
[:SENSe[1]]:VOLTage[:DC]:NPLCycles?
Model 2461 command
[:SENSe[1]]:<function>:NPLCycles
(on page 6-60), where
<function>
is
VOLTage[:DC]
,
CURRent[:DC]
, or
RESistance
.
[:SENSe[1]]:<function>:NPLCycles?
Notes
If you send
[:SENSe[1]]:NPLCycles
, it sets NPLCs for all functions.
Model 2400 command
[:SENSe[1]]:CURRent[:DC]:PROTection[:LEVel]
[:SENSe[1]]:CURRent[:DC]:PROTection[:LEVel]?
Model 2461 command
:SOURce[1]:<function>:<x>LIMit[:LEVel]
(on page 6-94)
Notes
Model 2400 command
[:SENSe[1]]:CURRent[:DC]:PROTection:RSYNchronize
[:SENSe[1]]:CURRent[:DC]:PROTection:RSYNchronize?
[:SENSe[1]]:VOLTage[:DC]:PROTection:RSYNchronize
[:SENSe[1]]:VOLTage[:DC]:PROTection:RSYNchronize?
Model 2461 command
Not available
Notes
Range synchronization is always turned on in Model 2461
Model 2400 command
[:SENSe[1]]:CURRent[:DC]:PROTection:TRIPped?
Model 2461 command
:SOURce[1]:<function>:<x>LIMit[:LEVel]:TRIPped?
(on page 6-95)
Notes
Model 2400 command
[:SENSe[1]]:CURRent[:DC]:RANGe:AUTO
[:SENSe[1]]:CURRent[:DC]:RANGe:AUTO?
[:SENSe[1]]:RESistance:RANGe:AUTO
[:SENSe[1]]:RESistance:RANGe:AUTO?
[:SENSe[1]]:VOLTage[:DC]:RANGe:AUTO
[:SENSe[1]]:VOLTage[:DC]:RANGe:AUTO?
Model 2461 command
[:SENSe[1]]:<function>:RANGe:AUTO
(on page 6-62), where
<function>
is
VOLTage[:DC]
,
CURRent[:DC]
, or
RESistance
.
[:SENSe[1]]:<function>:RANGe:AUTO?
Notes