Section 6: SCPI command reference
Model 2461 Interactive SourceMeter® Instrument Reference Manual
6-190
2461-901-01 A/November 2015
:TRIGger:BLOCk:BRANch:EVENt
This command branches to a specified block when a specified trigger event occurs.
Type
Affected by
Where saved
Default value
Command only
Recall settings
Instrument reset
Power cycle
Save settings
Not applicable
Usage
:TRIGger:BLOCk:BRANch:EVENt <blockNumber>, <event>, <branchToBlock>
<blockNumber>
The sequence of the block in the trigger model
<event>
The event that must occur before the trigger model branches the specified block
<branchToBlock>
The block number of the trigger model block to execute when the specified event
occurs
Details
The branch-on-event block goes to a branching block after a specified trigger event occurs. If the
trigger event has not yet occurred when the trigger model reaches the branch-on-event block, the
trigger model continues to execute the blocks in the normal sequence. After the trigger event occurs,
the next time the trigger model reaches the branch-on-event block, it goes to the branching block.
If you set the branch event to none, an error is generated when you run the trigger model.
The following table shows the constants for the events.
Trigger events
Event description
Event constant
No trigger event
NONE
Front-panel TRIGGER key press
DISPlay
Notify trigger block
<n>
(1 to 8); the trigger model generates a
trigger event when it executes the notify block
NOTify<n>
A command interface trigger:
•
Any remote interface: *
TRG
•
GPIB only: GET bus command
•
VXI-11: VXI-11 command
device_trigger
COMMand
Line edge (either rising, falling, or either based on the
configuration of the line) detected on digital input line
<n>
(1 to 6)
DIGio<n>
Line edge detected on TSP-Link synchronization line
<n>
(1 to 3)
TSPLink<n>
Appropriate LXI trigger packet is received on LAN trigger object
<n>
(1 to 8)
LAN<n>
Trigger event blender
<n>
(1 or 2), which combines trigger events
BLENder<n>
Trigger timer
<n>
(1 to 4) expired
TIMer<n>
Source limit condition occurs
SLIMit