Section 6: SCPI command reference
Model 2461 Interactive SourceMeter® Instrument Reference Manual
6-28
2461-901-01 A/November 2015
:CALCulate2:<function>:LIMit<Y>:CLEar[:IMMediate]
This command clears the results of the limit test defined by
Y
.
Type
Affected by
Where saved
Default value
Command only
Not applicable
Not applicable
Not applicable
Usage
:CALCulate2:<function>:LIMit<Y>:CLEar[:IMMediate]
<function>
The measure function:
•
Current:
CURRent[:DC]
•
Resistance:
RESistance
•
Voltage:
VOLTage[:DC]
The digitize function:
•
Digitize current:
DIGitize:CURRent
•
Digitize voltage:
DIGitize:VOLTage
<Y>
Limit number:
1
or
2
Details
Use this command to clear the test results of limit
Y
when the limit auto clear option is turned off. Both
the high and low test results are cleared.
To avoid the need to manually clear the test results for a limit, turn the auto clear option on.
If you send this command without the
<function>
parameter, it clears the limit for all measure
functions. It will not change the setting for a digitize function.
Example
:CALC2:VOLT:LIM1:CLE:AUTO OFF
:CALC2:VOLT:LIM1:AUD FAIL
:CALC2:VOLT:LIM1:LOW 0.25
:CALC2:VOLT:LIM1:UPP 2.5
:CALC2:VOLT:LIMIT1:STAT ON
:READ?
:CALC2:VOLT:LIMIT1:FAIL?
:CALC2:VOLT:LIM1:CLE
Set limit autoclear off.
Enable the beeper for limit 1 when a voltage
measurement exceeds the limit.
Set lower limit 1 for voltage to 0.25 V.
Set upper limit 1 for voltage to 2.5 V.
Enable limit 1 testing for voltage.
Make a reading; the limit is checked and results
display on the front panel
Return the test results; example output if the test
fails on the low limit:
LOW
Clear the test results.
Also see
:CALCulate2:<function>:LIMit<Y>:CLEar:AUTO
(on page 6-26)
:CALCulate2:<function>:LIMit<Y>:LOWer[:DATA]
(on page 6-30)
:CALCulate2:<function>:LIMit<Y>:UPPer[:DATA]
(on page 6-33)