Usage
serialno
= localnode.serialno
serialno
The serial number of the instrument
Details
This read-only attribute indicates the instrument serial number.
Example
display.clear()
display.settext(localnode.serialno)
Clears the unit's display.
Places the unit's serial number on the top line of its display.
Also see
(on page 7-134)
localnode.showerrors
This attribute sets whether or not the instrument automatically sends generated errors.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (RW)
Yes
Power cycle
Not saved
0 (disabled)
Usage
errorMode
= localnode.showerrors
localnode.showerrors =
errorMode
errorMode
Enables (1) or disables (0) show-errors state
Details
If this attribute is set to
1
, the instrument automatically sends any generated errors stored in the error queue, and
then clears the queue. Errors are processed after executing a command message (just before issuing a prompt,
if prompts are enabled).
If this attribute is set to
0
, errors are left in the error queue and must be explicitly read or cleared.
When using this command from a remote node,
localnode
should be replaced with the node reference, for
example
node[5].showerrors
.
Example
localnode.showerrors = 1
Enables sending of generated errors.
Also see