Section 7: Command reference
Model 2651A High Power System SourceMeter® Instrument Reference Manual
7-162
2651A-901-01 Rev. A / March 2011
Also see
None
scriptVar.save()
This function saves the script to nonvolatile memory or to a USB flash drive.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Function No
Usage
scriptVar
.save()
scriptVar
.save(
filename
)
scriptVar
The name of variable that references the script
filename
The file name to use when saving the script to a USB flash drive
Details
The
scriptVar.save()
function only saves a script to nonvolatile memory or a USB flash drive.
You are not required to qualify the path to the USB flash drive, but if you would like to, add
/usb1/
before the
fileName
(see Example 2).
You are not required to add the file extension, but if you would like to, the only allowed extension is
.tsp
(see
Example 2).
Example 1
test8.save()
Saves the script referenced by the variable
test8
to nonvolatile memory.
Example 2
test8.save("/usb1/myScript.tsp")
Saves the script referenced by the variable
test8
to a file named
myScript.tsp
on your
flash drive
Also see
scriptVar.source
This attribute holds the source code of a user script.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (RW)
(see
Details)
No
Not applicable
Not saved
Not applicable