Section 8: Troubleshooting guide
Model 2651A High Power System SourceMeter® Instrument Reference Manual
8-2
2651A-901-01 Rev. A / March 2011
Reading errors
When errors occur, the error messages are placed in the error queue (see
The following table lists commands associated with the error queue (see
(on
page 5-1) for more information):
Error queue command
Description
errorqueue.clear()
Clear error queue of all errors
errorqueue.count
Number of messages in the error/event queue
errorqueue.next()
Request error message
For example, the following commands request the next complete error information from the error
queue and then output the message portion of the error:
errorCode, message, severity, errorNode = errorqueue.next()
print(message)
Error summary list
Error summary
Error number
Error level
Error Message
-430 RECOVERABLE
Query
Deadlocked
-420 RECOVERABLE
Query
Unterminated
-410 RECOVERABLE
Query
Interrupted
-363 RECOVERABLE
Input
Buffer
Over-run
-350 RECOVERABLE
Queue
Overflow
-315
RECOVERABLE
Configuration Memory Lost
-314
RECOVERABLE
Save/ Recall Memory Lost
-292
RECOVERABLE
Referenced name does not exist
-286
RECOVERABLE
TSP Runtime error
-285 RECOVERABLE
Program
Syntax
-282
RECOVERABLE
Illegal program name
-281
RECOVERABLE
Cannot Create Program
-225
RECOVERABLE
Out of Memory or TSP Memory allocation error
-224
RECOVERABLE
Illegal Parameter Value
-223
RECOVERABLE
Too Much Data
-222
RECOVERABLE
Parameter Data Out of Range
-221 RECOVERABLE
Settings
Conflict
-220 RECOVERABLE
Parameter
-203 RECOVERABLE
Command
protected
-154
RECOVERABLE
String Too Long
-151
RECOVERABLE
Invalid String Data
-144
RECOVERABLE
Character Data Too Long
-141
RECOVERABLE
Invalid Character Data