Reading Status & Settings from the 2553/ 2553E
47
Example –
•
The user transmits: INTEGRATE?<NL>
The user reads the 2553/ 2553E and receives the characters: <SP>1<NL>
This indicates that the 2553/ 2553E is integrating results (i.e. Watt-Hour and related
results are being updated).
MEASURE?
The 2553/ 2553E places the ASCII character 1 (decimal 49) if it is presently performing
measurements, or the ASCII character 0 (decimal 48) if measurements are presently
frozen by the user. The MEASURE command controls this configuration.
Example –
•
The user transmits: MEASURE?<NL>
The user reads the 2553/ 2553E and receives the characters: <SP>0<NL>
This indicates that the 2553/ 2553E has frozen all measurement results.
SYNC?
The 2553/ 2553E places a single ASCII character representing the presently selected
source of synchronization for measurement results. The SYNC command controls this
configuration. The characters returned are as follows.
Character
Synchronization
0
Phase A Voltage Input
1
Phase A Current Input
2
Fixed 50Hz Synchronization
3
Fixed 60Hz Synchronization
4
Fixed 400Hz Synchronization
5
None
Table 29 – Data returned due to SYNC? command
Example –
•
The user transmits: SYNC?<NL>
The user reads the 2553/ 2553E and receives the characters: <SP>0<NL>
This indicates that the 2553/ 2553E is making measurements synchronized to the
Voltage Input signal.