52
2553/ 2553E Command Guide
Setting the Time Interval Between Bank Updates
This is achieved by the UPDATEn set of commands (n is a digit character between 0 and
4 inclusive, defining which bank is to have its update interval changed). Following reset,
all banks have an update interval of 250ms. In most applications this will suffice, thus
these commands are rarely used.
UPDATE0, UPDATE1, UPDATE2, UPDATE3 and UPDATE4
This command should be followed by an = character (decimal 61) and one or more digit
characters. The update interval for the bank controlled by the command is set to the
integer data number of 10ms intervals. Although allowed, setting the update rate to less
than 2 is not recommended.
Examples –
•
The user transmits: UPDATE0=25<NL>
The 2553/ 2553E sets the update interval for bank #0 to 250ms (25 x 10ms).
•
The user transmits: UPDATE1=1000<NL>
The 2553/ 2553E sets the update interval for bank #1 to 10 seconds (1000 x 10ms).
•
The user transmits: UPDATE0=25;UPDATE1=1000<NL>
The 2553/ 2553E sets the update interval for bank #0 to 250ms, and sets the update
interval for bank #1 to 10 seconds.
Selecting the Measurement Result(s) to Be Read
There are many thousands of possible measurement results which may be formatted into
a bank, thus the user needs to set which measurement results are desired, and the order in
which they should be presented, for each bank which the user will use.
This is achieved by the BANKn set of commands (n is a digit character between 0 and 4
inclusive, defining which bank is to have its definition changed). Following reset, all
banks are set to not be updated with any measurement results.
BANK0, BANK1, BANK2, BANK3 and BANK4
This command should be followed by a = character (decimal 61) and one or many (up to
50)
measurement result definitions. If many measurement result definitions are to be
given, then each should be separated by the / character (decimal 47). Alternatively, this
command may be followed by no data, in which case no measurement results will be
formatted into the specified bank.
The following paragraphs detail each available measurement result definition and their
allowable formats. A separate booklet is available from
Vitrek
that
contains the exact
method used to compute each of these measurement results.