2553/ 2553E Command Guide
Defining Historical Measurement Results
With the exception of HISTORY-SIZE, all of these return a 200 result set of results. The
first result formatted is the earliest point, the last is the latest. Results which have not
been placed into historical storage (i.e., at a time prior to starting the storage of historical
data) is returned as a zero or blank value (use HISTORY-SIZE to obtain the number of
valid data).
A-HISTORY[
phase
/PEAK]
The set of highest instantaneous absolute current amplitudes that have been stored
historically. Each result is floating point, in Amps, and is always positive. Note that this
yields 1600 characters in total. TOTAL is derived from the highest value for all phases.
A-HISTORY[
phase
/RMS]
The set of RMS current amplitudes that have been stored historically. Each result is
floating point, in Amps, and is always positive. Note that this yields 1600 characters in
total. TOTAL is the mean value for all phases.
A-HISTORY[
phase
/THD]
The set of evaluated current distortion results that have been stored historically. Each
result is floating point, in percent relative to fundamental, and is always positive. Note
that this yields 1600 characters in total. TOTAL is the mean value for all phases.
HISTORY-DATE
The set of dates, which correspond to each measurement result that has been stored
historically. Results that have not been placed into historical storage (i.e., at a time prior
to starting the storage of historical data) are returned as a string containing 11 <SP>
characters. Otherwise, each result is an ASCII string of format
mmm dd yyyy
(
mmm
is
the three letter month,
dd
is the numeric date,
yyyy
is the four digit year). Note that this
yields 2400 characters in total.
HISTORY-SIZE
The number of valid historically-stored sets of results that have presently been stored,
formatted as a three digit integer. This number can be in the range of zero to 200 (zero
indicating that no result sets have been stored and 200 indicating that all historical results
are valid). Note that this number of results will be found at the end of the set of results
obtained by any of the individual historically-stored data definitions.