TOS9213S/ TOS9213AS
7-19
DCW:RISETIME (D:RTIM)
Sets the voltage rise time (RISE TIME) for a DC withstanding voltage test. Also
inquires about the set or setting voltage rise time.
Message
Setting:
DCW:RISETIME_<
voltage rise time
>
D:RTIM_<
voltage rise time
>
Query:
DCW:RISETIME?
D:RTIM?
Program data
Data format: Real number
Set value:
0.1 to 200
Resolution:
0.1 for 0.1 to 99.9, 1 for 100 to 200
Unit:
s
(Example) To set the voltage rise time to 1 s,
D:RTIM 1
Response
To D:RTIM
?
, the current voltage rise time is returned.
(Example) If the voltage rise time is currently set to 2.0 s,
2.0
is returned.
DCW:DISCH (D:DIS)
Sets the discharge time (DISCH TIME), which is the duration of the forcible dis-
charge that occurs after DC withstanding voltage testing. Also inquires about the
current discharge time.
Message
Setting:
DCW:DISCH_<
DISCH TIME
>
D:DIS_<
DISCH TIME
>
Query:
DCW:DISCH?
D:DIS?
Program data
Data format: Real number
Set value:
0.0, 0.5 to 300
Resolution:
0.1 for 0.5 to 99.9, 1 for 100 to 300
Unit:
s
(Example) To set the discharge time to 1 s,
D:DIS 1
Response
To D:DIS
?
, the current discharge time is returned.
(Example) If the discharge time is currently set to 2.0 s,
2.0
is returned.