TOS9213S/ TOS9213AS
7-39
(Example) To set the FAIL mode to OFF,
FMOD OFF
Response
To FMOD
?
, the current FAIL mode setting is returned.
(Example) If the FAIL mode function is currently ON,
1
is returned.
DOUBLE ACTION (DAC)
Sets the double-action mode (DOUBLE ACTION). Also inquires about the current
double-action mode setting.
Message
Setting:
DOUBLEACTION_<{ON|OFF}>
DAC_<{ON|OFF}>
Query:
DOUBLEACTION?
DAC?
Program data
Data format: Character
Set value:
OFF (0), ON (1)
(Example) To set the double-action mode to OFF,
DAC OFF
Response
To DAC
?
, the current double-action mode setting is returned.
(Example) If the double-action mode is currently ON,
1
is returned.
SIGHVON (SHV)
Sets H.
V
ON of Status Signal Output. Also inquires about the H.
V
ON setting.
Message
Setting:
SIGHVON_<{ON|OFF}>
SHV_<{ON|OFF}>
Query:
SIGHVON?
SHV?
Program data
Data format: Character
Set value:
OFF (0), ON (1)
(Example) To set H.
V
ON of Status Signal Output to OFF,
SHV OFF
Response
To SH
V?
, the current H.
V
ON setting for Status Signal Output is returned.
(Example) If H.
V
ON for Status Signal Output is currently set to ON,
1
is returned.