TOS9213S/ TOS9213AS
7-13
Response
To ERR
?
, the contents of the error register are returned.
(Example) When Bit 3 of the error register is set,
8
is returned.
FAIL
Inquires about the contents of the fail register.
For details on the fail register, see "7.5 Registers".
Message
Query:
FAIL?
Response
To FAIL
?
, the contents of the fail register are returned.
(Example) When Bit 4 of the fail register is set,
16
is returned.
FUNCTION (FUN)
Shifts the LCD to each mode (setting screen) and Inquires about the current mode
(displayed screen).
Message
Setting:
FUNCTION_<
value
>
FUN_<
value
>
Query:
FUNCTION?
FUN?
Program data
Data format: Integer
Set value:
1: DCW; 2: IR; 3: AUTO TEST; 4: AUTO EDIT;
5: SYSTEM; 6: OFFSET ADJ; 7: INTERFACE
(Example) To change the contents of the program (to shift to AUTO EDIT)
FUN 4
Response message
To FUN
?
, the current mode is returned.
(Example) If the current mode is the DC withstanding voltage test,
0
is returned.
INVALID (INV)
Used to inquires about the contents of the invalid-setting register.
For details on the invalid-setting register, see "7.5 Registers".
Message
Query:
INVALID?
INV?
Response
To IN
V?
, the contents of the invalid-setting register are returned.
(Example) When Bit 3 of the invalid-setting register is set,
8
is returned.