Operating Instructions - CPC8000
Table 8.2 - Mensor Command Set
Command/Query
Data
Function/Response
?
See section 8.2 text on previ-
ous page
Returns data per the current output format.
Acquire?
15 char string.
Ex:
Acquire? Test_stand_1
Returns:
<sp>(YES or NO), CCC…
CCCcr><lf>
This command is used when multiple
computers would like to control the instrument.
Yes if acquisition is successful.
No if instrument is being controlled with
another computer.
CCC… = name of controlling computer.
See: Release? and Unlock
Actualpress
<sp><n>,Decimal value
Sets the actual/measured pressure for
<n>th linearity point (0 based index). Must
be within 1% of “Desiredpress” value.
See Section 7.5.2.5.3, Linearize.
Actualpress?
<sp><n>
<sp>+n.nn<cr><lf>
Returns the actual/measured pressure for
<n>th linearity point (0 based index).
See Section 7.5.2.5.3, Linearize.
Address
0-31
Sets the GPIB Address.
Address?
<sp> nn <cr><lf>
Returns the GPIB Address.
Asset_tag
16 char string
General purpose string for customer use.
Asset_tag?
<sp>ssssssssssssssss<cr><lf>
Return customer asset tag string.
Autorange
ON or OFF
Sets whether the autorange function is
enabled or disabled.
Autorange?
<sp>(ON or OFF)<cr><lf>
Returns whether the autorange function is
enabled or disabled.
Autovent
<ON,OFF,YES,NO,1,0>
The autovent function performs a controlled
vent of the system when a setpt of 0 is sent
and the controller is in gauge mode.
Autovent?
<sp>(1 or 0)<cr><lf>
Returns whether autovent is enabled or
disabled.
Autozero
none
Re-zero all the ranges. These adjust-
ments are not password protected and
are not saved through power cycles. This
command takes approximately 60 seconds
but may take longer depending on the time
to become stable.
Autozero?
S,T,X,X
Returns autozero data where S represents
state (0 = complete, 1 = local autozero, 2 =
remote autozero), T represents the estimat-
ed time to complete in seconds, and x is a
(0) character since this data location is not
used at this time.
Autozeroabort
none
Aborts autozero. Any sensors that have
been zeroed will not revert to previous zero
ofsets.
Summary of Contents for CPC8000
Page 6: ...High End Pressure Controller CPC8000 6 Operating Instructions CPC8000 NOTES ...
Page 10: ...High End Pressure Controller CPC8000 10 Operating Instructions CPC8000 NOTES ...
Page 24: ...High End Pressure Controller CPC8000 24 Operating Instructions CPC8000 NOTES ...
Page 84: ...High End Pressure Controller CPC8000 84 Operating Instructions CPC8000 Notes ...
Page 112: ...High End Pressure Controller CPC8000 Operating Instructions CPC8000 113 NOTES ...