High-End Pressure Controller
CPC8000
Operating Instructions - CPC8000
87
Baro?
<sp>+n.nn<cr><lf>
Returns reading from barometric sensor or
“NO BAROMETER” if one isn’t installed.
Barocaldisable
YES,NO
Not used, kept for backwards compatibility.
Barocaldisable?
<sp>(YES or NO)<cr><lf>
Not used, kept for backwards compatibility.
Calculate_as_
found_linearity
Calculate linearity slopes and inter-
cepts from “Desiredpress”/”Actualpress”
pressures.
Caldisable
YES,NO
Sets whether or not calibration of the active
sensor is disabled.
Caldisable?
<sp>(YES or NO)<cr><lf>
Returns whether or not calibration of the
active sensor is disabled.
Cerr
none
Clears the error queue.
Chanfunc
Press, peak, rate, rate setpt,
dio, units
Sets the secondary display function mode,
works identical to Chanfunc2 (kept for
backwards compatibility).
Chanfunc?
<sp>CCCCC…<cr><lf>
Returns the secondary alternate function
mode, works identical to Chanfunc2 (kept
for backwards compatibility).
Chanfunc2
Press, peak rate, rate setpt,
dio, units
Sets the secondary display function mode.
Chanfunc2?
<sp>CCCCC…<cr><lf>
Returns the secondary alternate function
mode.
Chanfunc3
Press, peak rate, rate setpt,
dio, units
Sets the tertiary display function mode.
Chanfunc3?
<sp>CCCCC…<cr><lf>
Returns the tertiary alternate function mode.
Cmdset
Mensor, SCPI
Activates remote command set for instru-
ment emulation modes.
Cmdset?
<sp><CCCCCC><cr><lf>
Returns active command set identifer.
Control
Instrument placed in Control Mode
Control?
<sp>(YES or NO)<cr><lf>
Returns YES if instrument is in control. No if
otherwise.
Crate
Slow, Medium, Fast, Variable
Sets the control rate. Variable mode is a
predetermined user-defned rate setpt.
Crate?
<sp>CCCCCC<cr><lf>
Returns the control rate – CCCC is variable
in length and corresponds to the param-
eters for the CRATE command.
Ctype?
Returns the type of regulator.
Decpt?
<sp>n<cr><lf>
Returns the number of decimal points.
(See: Resolution)
Default
none
Sets the default values.
Desiredpress
<sp><n>,Decimal value
Sets the reference pressure for <n>th linear-
ity point (0 based index). Must be greater
than <n-1>’s “Desiredpress” value.
Summary of Contents for CPC8000
Page 6: ...High End Pressure Controller CPC8000 6 Operating Instructions CPC8000 NOTES ...
Page 10: ...High End Pressure Controller CPC8000 10 Operating Instructions CPC8000 NOTES ...
Page 24: ...High End Pressure Controller CPC8000 24 Operating Instructions CPC8000 NOTES ...
Page 84: ...High End Pressure Controller CPC8000 84 Operating Instructions CPC8000 Notes ...
Page 112: ...High End Pressure Controller CPC8000 Operating Instructions CPC8000 113 NOTES ...