High-End Pressure Controller
CPC8000
88
Operating Instructions - CPC8000
Desiredpress?
<sp><n>
<sp>+n.nn<cr><lf>
Returns the reference pressure for <n>th
linearity point (0 based index).
See Section 7.5.2.5.3, Linearize.
DHCP
ON or OFF
If no DHCP server is found when DHCP is
turned on, DHCP will be turned of.
DHCP?
<sp>(YES or NO)<cr><lf>
Returns current status of DHCP.
DIO
2 or 0
2 turns on the frst digital output pin, 0 turns
it of.
DIO?
<sp>n<cr><lf>
Returns status of the frst input and output
pin. Bit0 = input’s status, Bit1 = output’s
status.
DOUTFUNC
<n><sp><CCCCCC><cr><lf>
Sets the function for output pin <n> to
NONE, STABLE, PUMP, MEASURE,
CONTROL, or VENT.
DOUTFUNC? <n>
<sp><CCCCCC><cr><lf>
Returns the function for output pin <n>.
DINFUNC
<n><sp><CCCCCC>
Sets the function for input pin <n> to NONE,
MEASURE, CONTROL, VENT, KEYLOCK,
or START.
DINFUNC? <n>
<sp><CCCCCC><cr><lf>
Returns the function for output pin <n>.
DOUTSTATE
<n><sp>HIGH/LOW/1/0>
Turns output pin <n> to high or low. Sets the
function for that pin to “None”.
DIOSTATE?
<sp><n><cr><lf>
Bit 0-7 is the status of the input bits. Bit 8-15
are the status of the output bits. Returns an
integer between 0 and 65536.
DOC
mm/dd/yyyy
Sets the date of cal for the active sensor.
DOC?
<sp>mm/dd/yyyy<cr><lf>
Returns the date of cal for the active sensor.
DOM?
<sp> mm/dd/yyyy<cr><lf>
Returns the date of manufacture.
Error?
<sp> text description <cr><lf>
Returns the next error in the error queue.
Errorno?
<sp>Enn-text<cr><lf>
Returns pcs400 error code and text.
Filter
Of, Low, Normal, High
Sets the reading flter 0, 80%, 92%, 95%.
Filter?
<sp> (flter)<cr><lf>
Returns the reading flter.
Gasdensity
Value in lb/cuft or “NITRO-
GEN” or “DRY AIR”
Sets the head pressure gas density in lb/cuft.
Gasdensity?
<sp>+n.nn<cr><lf>
Gets the head pressure gas density in lb/cuft.
Gastemp
Value in degrees F
Sets the head pressure gas temperature.
Gastemp?
<sp>+n.nn<cr><lf>
Gets the head pressure gas temperature.
Gateway
nnn.nnn.nnn.nnn
Sets the Ethernet gateway address.
Gateway?
<sp>nnn.nnn.nnn.
nnn<cr><lf>
Gets the Ethernet gateway address.
Height
value in inches
Sets the head pressure height in inches.
Height?
<sp>+n.nn<cr><lf>
Gets the head pressure height in inches.
Id?
<sp>MENSOR, CPC8000,
ssssss,v.v.vv<cr><lf>
Ssssss is the serial number, v.v.vv is the
CPC8000 software version.
Summary of Contents for CPC8000
Page 6: ...High End Pressure Controller CPC8000 6 Operating Instructions CPC8000 NOTES ...
Page 10: ...High End Pressure Controller CPC8000 10 Operating Instructions CPC8000 NOTES ...
Page 24: ...High End Pressure Controller CPC8000 24 Operating Instructions CPC8000 NOTES ...
Page 84: ...High End Pressure Controller CPC8000 84 Operating Instructions CPC8000 Notes ...
Page 112: ...High End Pressure Controller CPC8000 Operating Instructions CPC8000 113 NOTES ...