Variable
Function
Set M1
3
Marker 1
3 is set.
12.2.4.2
Output area
Data transmission from the PR 5410 to the master is via the output area.
The weight or data information requested by the master is transmitted to the master by
PR 5410.
The PR 5410 has write access, the master has read access.
Byte
Name
Description
0
Read_Value (MSB)
e.g., gross value
1
Read_Value (MSB)
e.g., gross value
2
Read_Value (MSB)
e.g., gross value
3
Read_Value (LSB)
e.g., gross value
4
Read_Value_Selected
e.g. gross
5
Write
Active
Power
Fail
Out 3
Out 2
Out 1
Limit 3 Limit 2 Limit 1 Status
6
Cmd
Busy
Cmd
Error
Inp. 3
Inp. 2
Inp. 1
Tare
Active
Cal Ac-
tive
Test
Active
Command status
7
Dim-
med
Stand-
still
Inside
ZSR
Center
Zero
Below
Zero
Over-
load
Above
Max
ADC
error
Device status
Bit 7
Bit 6
Bit 5
Bit 4
Bit 3
Bit 2
Bit 1
Bit 0
Variable
Function
Read_Value
The weight value is transmitted as a binary 32-bit value with plus or minus sign.
Type of data: DINT
Read_Value_Selected
Acknowledgment of the transmitted value.
Write_Active
The function selected with
Write_Value_Select
is executed once.
This bit is deleted if
Write_Value_Select
is set to 0.
Power_Fail
Is set when switching on the device.
Is reset by
ResPower
with transition from 0→1.
Cmd_Busy
The device is busy executing a command (e.g., the device has received a tare com-
mand and is waiting for standstill).
Cmd_Error
The device has interrupted the execution of a command (e.g.
standstill
could not
be reached within the defined standstill time).
The error number can be read at
Last error
. It is set only if an action is executed.
Tare_Active
The scale has been tared.
Cal_Active
The device is/was configured.
When this bit is 1, the scale parameters (EXPO/UNIT/STEP) must be read again.
Set after "Power on" and reset after reading the FSD.
X3 Process Indicator PR 5410
12 Fieldbus interface
Minebea Intec
EN-296