Bit
Description
50 = scale interval "50"
LASTERROR
Last error byte; see also
CmdError
bit, number of "last error":
31 = no standstill was reached (e.g. when taring).
33 = negative weight value when taring and legal-for-trade mode
on.
47 = no zero setting; weight not within zero setting range.
107 = no standstill with
Getfixtare
.
12.2.6.6
Register 5: Device type and software release (Read)
Bit 7
Bit 6
Bit 5
Bit 4
Bit 3
Bit 2
Bit 1
Bit 0
Byte 0
TYPE MSB
Byte 1
TYPE LSB
Byte 2
MAIN VERSION
Byte 3
SUBVERSION
12.2.6.7
Register 6: Board number (Read)
Bit 7
Bit 6
Bit 5
Bit 4
Bit 3
Bit 2
Bit 1
Bit 0
Byte 0
board number MSB
Byte 1
board number MSB
Byte 2
board number MSB
Byte 3
board number LSB
e.g.: 148388723 = 08D83B73
hex
12.2.6.8
Register 7: reserved
12.2.6.9
Registers 8…22: Weight data (Read)
The gross, net, and tare weight are stored as DINT fix point. The real data value is derived
from DINT and EXPO as follows:
Value
Real
= reading
DINT
× 10
(-EXPO)
Register 8
Current gross value
Register 9
Current net value, if tared; otherwise gross
Register 10
Current tare value, if tared, otherwise 0
Register 11
Value in the front readout
Register 12
Reserved
Register 13
Reserved
Register 14
Max (FSD)
12 Fieldbus interface
X3 Process Indicator PR 5410
EN-303
Minebea Intec