Byte 3
TareActive CalActive
TestActive
Note:
Byte 0 corresponds to byte 7 in the output area; for weight error, see also the table in
Chapter
Bit
Description
ADUERR
Error of analog conversion/load cell circuit (OR function of the E1,
E3, E7 bits).
>Max
Gross value is higher than Max (FSD); the scale range was exceeded.
OVL
Scale overload; Gross >Max + overload,
Error 2
BELOW0
Gross weight is negative (<0-¼ d).
CZERO
"Center of zero", weight within 0 ±¼ d.
INZSR
The gross weight is within the zero setting range.
STND
The scale is stable.
OUT
Below zero or above Max (FSD) (0-¼ d >weight value >Max).
E7
The measuring signal is negative (inverse conversion),
Error 7
E6
No or too low sense voltage;
Error 6
E3
The measuring signal is >36 mV (no end of conversion),
Error 3
E1
Arithmetic error (overflow),
Error 1
E9
No communication with xBPI scale,
Error 9
CmdError
Error during execution (cmdError); e.g. the "taring" operation is not
handled, because the scale is not stable. The error is stored
in
LastError
(register 4).
The bit is reset with the
ResetError
bit (register 121).
ActionActive
The operation is being processed; processing is ongoing.
PowerFail
Power failure; is always set after power on. The
PowerFail
bit is reset
with the
ResetPWF
bit (register 85) "Reset power failure."
TestActive
The analog test is busy.
CalActive
The device is/was configured.
When this bit is 1, the scale parameters (EXPO/UNIT/STEP) must be
read again.
Set after "Power on" and reset after reading the FSD.
TareActive
The instrument was tared.
12.2.6.3
Register 2: State of state-controlled action bits (Read)
Signal state
Bit 7
Bit 6
Bit 5
Bit 4
Bit 3
Bit 2
Bit 1
Bit 0
12 Fieldbus interface
X3 Process Indicator PR 5410
EN-301
Minebea Intec