Remote Control Commands
R&S
®
NGA100
98
User Manual 5601.8919.02 ─ 03
MAX
Set the OPP value to maximum
Default unit: W
Example:
INST OUT1
POW:PROT:LEV 5
Set the OPP value of channel 1 to 5 W.
POW:PROT:LEV?
-> 5
Return the selected channel OPP value.
[SOURce:]POWer:PROTection:TRIPped?
This command queries the OPP tripped status of the previous selected channel
Return values:
<arg0>
1 | 0
1
OPP is tripped.
0
OPP is not tripped.
Example:
INST OUT1
POW:PROT:TRIP?
-> 0
OPP state of Ch 1 has not tripped.
Usage:
Query only
[SOURce:]POWer:PROTection[:STATe]
<arg0>
This command defines or queries the OPP state for the previous selected channel.
Parameters:
<arg0>
ON | OFF | 1 | 0
ON | 1
OPP is activated.
OFF | 0
OPP is deactivated.
Example:
INST OUT1
POW:PROT ON
OPP of Ch 1 is activated.
POW:PROT?
Return the OPP state of the selected channel.
[SOURce:]POWer:PROTection:CLEar
This command resets the OPP state of the selected channel. If the OPP has tripped
the red blinking OPP message on the display will be cleared for the selected channel.
Usage:
Event
Configuration Commands