Remote Control Commands
R&S
®
NGA100
83
User Manual 5601.8919.02 ─ 03
7.3.3
Voltage Setting
The
SOURce:VOLTage
subsystem contains the commands for setting the voltage of
the output channels. The default unit is V.
Example: Configuring the output voltage
This example contains all commands to configure and query the output voltage.
// ************************************************
// Select the channel
// ************************************************
INST OUT1
// ************************************************
// Set the voltage value
// ************************************************
// selects a channel and sets the voltage
VOLT 10
// queries the output voltage of a channel
VOLT?
// response: 10.000
// sets the voltage to maximum or minimum respectively
VOLT MAX
VOLT MIN
// ************************************************
// Query the range of the voltage values
// ************************************************
// queries the upper and lower limit of the output voltage
VOLT? MIN
// response: 0.000
VOLT? MAX
// response: 35.05 for NGA101, NGA102 and 100.05 for NGA141, NGA142
[SOURce:]VOLTage[:LEVel][:IMMediate][:AMPLitude]
<Voltage>
This command defines or queries the voltage value of the selected channel.
Parameters:
<Voltage>
{<Voltage> | MINimum | MAXimum | DEFault}
<Voltage>
0.0 V to 35.05 V for NGA101 and NGA102 models (adjustable in
1 mV steps)
0.0 V to 100.05 V for NGA141 and NGA142 models (adjustable
in 10 mV steps)
MINimum
0.000E+00
MAXimum
35.05 V for NGA101 and NGA102 models
100.05 V for NGA141 and NGA142 models
Default unit: V
Configuration Commands