118165-001 REV. A
475 Wireless Boulevard • Hauppauge, New York 11788, USA • www.spellmanhv.com • T:+1 631.630.3000 • F:+1 631.435.1620
40
4.4.14
Request Exposure Time Set-point
Description:
The host requests that the firmware report the current exposure
time set point. This is the current value of the exposure control
timer, not the actual time of the last exposure. The value is 0-12000
in mSec. Note: This command reads the value currently being used
by the system, which is mode dependent. In 2 point mode this
value is calculated using the mAs value along with the tube table. In
manual mode this value is set using the “Set Exposure Time”
command. When in 2 point mode the value read by this command
is not the same as what is set using the “Set Exposure Time”
Command.
Direction:
Host to supply
Syntax:
<STX><52><,><CSUM><ETX>
Response:
<STX><52><,><ARG><,><CSUM><ETX>
Where:
<ARG> = 0 - 12000 in ASCII format, scaling = msec/Bit
Example:
<STX>52,<CSUM><ETX>