Programming Manual MSO/UPO2000 Series
68 / 156
Instruments.uni-trend.com
Query returns {RUN|STOP}.
For example
:
:PF:OPERate
RUN
Run Pass/Fail test.
:PF:OPERate
?
Query returns RUN.
:PF:OUTPut
Command format
:
:PF:OUTPut
{PASS|FAILED}
:PF:OUTPut?
Functional description
:
This command is used to set or query the output of Pass/Fail test.
Return format
:
Query returns {PASS|FAILED}.
For example
:
:PF:OUTPut
PASS
Set the output of Pass/Fail test to PASS.
:PF:OUTPut
?
Query returns PASS.
:PF:STOP:TYPe
Command format
:
:PF:STOP:TYPe
{PCOUNT|FCOUNT}
:PF:STOP:TYPe?
Functional description
:
This command is used to set or query stop type of Pass/Fail test.
PCOUNT represents the number of pass; FCOUNT represents the number of fail
Return format
:
Query returns {PCOUNT|FCOUNT}.
For example
:
:PF:STOP:TYPe
PCOUNT
Set stop type of Pass/Fail test to PCOUNT..
:PF:STOP:TYPe
?
Query returns PCOUNT
:PF:STOP
:QUALifier
Command format
:
:PF:STOP
:QUALifier {LEQual | GEQual}
:PF:STOP
:QUALifier
?
Functional description
:
This command is used to set or query the stop condition of Pass/Fail test.
GEQual represents greater than or equal to; LEQual represents less than or equal to.
Return format
:
Query returns {LEQual | GEQual}.
For example
:
:PF:STOP
:QUALifier GEQual
Set the stop condition of Pass/Fail test to ≥.
:PF:STOP
:QUALifier?
Query returns GEQual.