95x Series Operating Manual - May 17, 2022
Page
143
of
155
Example
–
ADD,DCI,5.0,0.0,0.0,50e-6,,,ABORT
Configures the following (in order)
–
A DCI type test step
5 second test time
0 second delay time
No minimum leakage current
50uA maximum leakage current
No arc detection (so following field is also <EMPTY>)
Abort test sequence if fails
PULSE CONFIGURATION FIELDS
Field # Field Format Value
1
<STRING>
PULSE
2
<NR3>
Level (in V)
3
<STRING>
POSITIVE - Unve
NEGATIVE
–
Unipolar
–
ve
BIPOLAR - Bipolar
4
<NR3>
Ramp time (tr) (in seconds)
5
<NR3>
Hold Time (th) (in seconds)
6
<NR3>
Breakdown Current Limit (in A)
7
<NR1>
Time period for arc detection (4, 10, 15, 20, 30 or 40)
<EMPTY>
Arc detection disabled
8
<NR1>
Arc detection limit (in mA)
<EMPTY>
Only valid if arc detection disabled in field 7
9
<STRING>
ABORT - Abort on failure
CONT - Continue on failure
Example
–
ADD,PULSE,250.0,POSITIVE,1e-3,10e-3,0.05,4,10,ABORT
Configures the following (in order)
–
A PULSE type test step
A single positive pulse
1ms ramp time
10ms hold time
50mA maximum current
4us arc detection
10mA arc detection limit
Abort test sequence if fails
BRKDN CONFIGURATION FIELDS
Field # Field Format Value
1
<STRING>
BRKDN