95x Series Operating Manual - May 17, 2022
Page
140
of
155
RAMP - Discharge same as ramp
11
<STRING>
ABORT - Abort on failure
CONT - Continue on failure
Example
–
ADD,GB,25.0,60.0,5.0,0.0,60.0,RMSO,0.0,0.1,FAST,ABORT
Configures the following (in order)
–
A GB type test step
25A test current
60Hz test frequency
5V max open circuit voltage
0 second ramp time
60 second dwell time
Check RMS impedance
No minimum impedance
0.1
Ω
maximum impedance
Fast discharge
Abort test sequence if fails
LOW
Ω
CONFIGURATION FIELDS
Field # Field Format Value
1
<STRING>
LOWOHM
2
<BOOL>
1 or Y - Test using 2-wire mode
0 or N - Test using 4-wire mode
3
<NR3>
Test time (in seconds)
<EMPTY>
Step is to be user terminated
4
<NR3>
Check delay (in seconds)
5
<NR3>
Minimum Resistance Limit (in
Ω
)
6
<NR3>
Maximum Resistance Limit (in
Ω
)
7
<STRING>
ABORT - Abort on failure
CONT - Continue on failure
Example
–
ADD,LOWOHM,0,5.0,0.0,1.25,1.75,ABORT
Configures the following (in order)
–
A Low
Ω
type test step
Use 4-wire mode
5 second test time
0 second delay
1.25
Ω
minimum resistance
1.75
Ω
maximum resistance
Abort test sequence if fails