SCPI Command Reference
17-63
:ADVanced:NTOLerance <NRf>
[:SENSe[1]]:AVERage:ADVanced:NTOLerance <NRf>
Set moving
fi
lter noise
fi
lter
Parameters
<NRf> =
0 to 105
Specify
fi
lter noise
fi
lter in %
Query
:NTOLerance?
Query
fi
lter noise
fi
lter value
Description
When the advanced
fi
lter is enabled, using the next command, a noise win-
dow is used with the moving
fi
lter. This command is used to specify the
noise window. If readings are within the noise window, the moving
fi
lter
operates normally. If, however, a reading falls outside the window, the
buffer is
fl
ushed of old readings and
fi
lled with the new reading.
:ADVanced[:STATe] <b>
[:SENSe[1]]:AVERage:ADVanced[:STATe] <b>
Enable/disable advanced
fi
lter
Parameters
<b> =
0 or OFF
Disable advanced
fi
lter
1 or ON
Enable advanced
fi
lter
Query
[:STATe]?
Query state of advanced
fi
lter
Description
This command is used to enable or disable the advanced
fi
lter. When
enabled, the noise window is used with the moving
fi
lter. When disabled,
the noise window is not used.
Summary of Contents for 6430
Page 26: ......
Page 32: ......
Page 78: ...2 14 Connections ...
Page 98: ...3 20 Basic Source Measure Operation ...
Page 138: ...5 30 Source Measure Concepts ...
Page 156: ...6 18 Range Digits Speed and Filters ...
Page 168: ...7 12 Relative and Math ...
Page 176: ...8 8 Data Store ...
Page 202: ...9 26 Sweep Operation ...
Page 248: ...11 22 Limit Testing ...
Page 310: ...16 6 SCPI Signal Oriented Measurement Commands ...
Page 418: ...17 108 SCPI Command Reference ...
Page 450: ...18 32 Performance Verification ...
Page 477: ...A Specifications ...
Page 489: ...B StatusandErrorMessages ...
Page 498: ...B 10 Status and Error Messages ...
Page 499: ...C DataFlow ...
Page 503: ...D IEEE 488BusOverview ...
Page 518: ...D 16 IEEE 488 Bus Overview ...
Page 519: ...E IEEE 488andSCPI ConformanceInformation ...
Page 523: ...F MeasurementConsiderations ...
Page 539: ...G GPIB488 1Protocol ...
Page 557: ......