SCPI Command Reference
17-77
:POINts?
:SOURce[1]:LIST:CURRent:POINts?
Query length of I-Source list
:SOURce[1]:LIST:VOLTage:POINts?
Query length of V-Source list
Description
This command is used to determine the length of the speci
fi
ed source list.
The response message indicates the number of source values in the list.
Con
fi
gure memory sweep
A memory sweep lets you perform a sweep using setups stored in up to 100 memory loca-
tions. This allows you to sweep using multiple source-measure operations and math expres-
sions. See
CALCulate1
.
NOTE
In order to run a memory sweep, the memory function must be selected, and the trig-
ger count must be the same as the number of memory points in the sweep. Use the
:FUNCtion:MODE command to select the MEMory function. See “Select function
mode.” Arm count (ARM:COUNt) and trigger count (TRIGger:COUNt) are set from
the Trigger Subsystem.
When memory is initialized (:SYSTem:MEMory:INITialize), all 100 memory loca-
tions for a memory sweep are initialized to the present setup configuration of the
SourceMeter with CALC 1 disabled. User-defined math expressions are replaced
with the “Power” math expression.
Error 809 “Source memory location revised” occurs when a memory sweep refer-
ences an expression that no longer exists. The memory sweep is revised to disable
CALC1.
In order to execute a memory sweep, the product of the arm count and trigger count
should be at least the same as the number of points in the sweep.
Use the :FUNCtion:MODE command to select the MEMory function. See
Select function
mode
. Arm count (ARM:COUNt) and trigger count (TRIGger:COUNt) are set from the Trigger
Subsystem.
:SAVE <NRf>
:SOURce[1]:MEMory:SAVE <NRf>
Save setup in speci
fi
ed memory location
Parameters
<NRf> = 1 to 100
Specify memory location
Description
This command is used to save the present instrument setup in a memory
location. Up to 100 setups can be saved. The following settings are saved in
each source memory location:
SENSe[1]:CURRent:NPLCycles
SENSe[1]:RESistance:NPLCycles
SENSe[1]:VOLTage:NPLCycles
SENSe[1]:FUNCtion:CONCurrent
SENSe[1]:FUNCtion:ON
SENSe[1]:FUNCtion:OFF
Summary of Contents for 6430
Page 26: ......
Page 32: ......
Page 78: ...2 14 Connections ...
Page 98: ...3 20 Basic Source Measure Operation ...
Page 138: ...5 30 Source Measure Concepts ...
Page 156: ...6 18 Range Digits Speed and Filters ...
Page 168: ...7 12 Relative and Math ...
Page 176: ...8 8 Data Store ...
Page 202: ...9 26 Sweep Operation ...
Page 248: ...11 22 Limit Testing ...
Page 310: ...16 6 SCPI Signal Oriented Measurement Commands ...
Page 418: ...17 108 SCPI Command Reference ...
Page 450: ...18 32 Performance Verification ...
Page 477: ...A Specifications ...
Page 489: ...B StatusandErrorMessages ...
Page 498: ...B 10 Status and Error Messages ...
Page 499: ...C DataFlow ...
Page 503: ...D IEEE 488BusOverview ...
Page 518: ...D 16 IEEE 488 Bus Overview ...
Page 519: ...E IEEE 488andSCPI ConformanceInformation ...
Page 523: ...F MeasurementConsiderations ...
Page 539: ...G GPIB488 1Protocol ...
Page 557: ......