gure voltage and current sweeps
There are two methods to con
fi
gure the start and stop levels of a sweep. You can use either
the :STARt and :STOP commands or you can use the :CENTer and :SPAN commands.
NOTE
In order to run a sweep, the selected source must be in the sweep sourcing mode and
the trigger count should be the same as the number of source-measure points in the
sweep.
Use the :FUNCtion:MODE command to select the current or voltage source function. See
Select function mode
. Use the :CURRent:MODE or VOLTage:MODE command to select the
SWEep sourcing mode. See
Select sourcing mode
. The trigger count is set using the
TRIGger:COUNt command. See
Trigger subsystem
.
:RANGing <name>
:SOURce[1]:SWEep:RANGing <name>
Select source ranging mode
Parameters
<name> = BEST
Use the best
fi
xed mode
AUTO
Use the most sensitive source range for
each sweep level
FIXed
Use the present source range for the
entire sweep
Query
:RANGing?
Query source ranging mode
Description
This command is used to select the source ranging mode for sweeps. With
BEST selected, the SourceMeter will select a single
fi
xed source range that
will accommodate all the source levels in the sweep. For front panel opera-
tion, this is the BEST FIXED option.
With AUTO selected, the SourceMeter will automatically go to the most
sensitive source range for each source level in the sweep. For front panel
operation, this is the AUTO RANGE option.
With FIXed selected, the source remains on the range that it is presently on
when the sweep is started. For sweep points that exceed the source range
capability, the source will output the maximum level for that range. For
front panel operation, this is the FIXED option.
:SPACing <name>
:SOURce[1]:SWEep:SPACing <name>
Select scale for sweep
Parameters
<name> = LINear
Linear scale
LOGarithmic
Logarithmic scale
Query
:SPACing?
Query scale for sweep
Description
This command is used to select the scale for the sweep. With LINear
selected, the source-measure points in the sweep will be performed on a lin-
ear scale. With LOGarithmic selected, the source-measure points will be
performed on a logarithmic scale.
Summary of Contents for 6430
Page 26: ......
Page 32: ......
Page 78: ...2 14 Connections ...
Page 98: ...3 20 Basic Source Measure Operation ...
Page 138: ...5 30 Source Measure Concepts ...
Page 156: ...6 18 Range Digits Speed and Filters ...
Page 168: ...7 12 Relative and Math ...
Page 176: ...8 8 Data Store ...
Page 202: ...9 26 Sweep Operation ...
Page 248: ...11 22 Limit Testing ...
Page 310: ...16 6 SCPI Signal Oriented Measurement Commands ...
Page 418: ...17 108 SCPI Command Reference ...
Page 450: ...18 32 Performance Verification ...
Page 477: ...A Specifications ...
Page 489: ...B StatusandErrorMessages ...
Page 498: ...B 10 Status and Error Messages ...
Page 499: ...C DataFlow ...
Page 503: ...D IEEE 488BusOverview ...
Page 518: ...D 16 IEEE 488 Bus Overview ...
Page 519: ...E IEEE 488andSCPI ConformanceInformation ...
Page 523: ...F MeasurementConsiderations ...
Page 539: ...G GPIB488 1Protocol ...
Page 557: ......