17-66
SCPI Command Reference
Select range
:RANGe <n>
:SOURce[1]:CURRent:RANGe <n>
Select range for I-Source
:SOURce[1]:VOLTage:RANGe <n>
Select range for V-Source
Parameters
<n> =
-105e-3 to 105e-3
Specify I-Source level (amps)
-210 to 210
Specify V-Source level (volts)
DEFault
I-Source: 100µA range
V-Source: 20V range
Minimum
I-Source: 1pA range (Remote PreAmp)
1µA range (Mainframe-only)
V-Source: 200mV range
MAXimum
I-Source: 100mA range
V-Source: 200V range
UP
Select next higher range
DOWN
Select next lower range
Query
:RANGe?
Query range for speci
fi
ed source
:RANGe? DEFault
Query *RST default source range
:RANGe? MINimum
Query lowest source range
:RANGe? MAXimum
Query highest source range
Description
This command is used to manually select the range for the speci
fi
ed source.
Range is selected by specifying the approximate source magnitude that you
will be using. The instrument will then go to the lowest range that can
accommodate that level. For example, if you expect to source levels around
3V, send the following command:
:SOURce:VOLTage:RANGe 3
The above command will select the 20V range for the V-Source.
As listed in
Parameters
, you can also use the MINimum, MAXimum and
DEFault parameters to manually select the source range. The UP parameter
selects the next higher source range, while DOWN selects the next lower
source range.
Note that source range can be selected automatically by the instrument. See
the next command.
Summary of Contents for 6430
Page 26: ......
Page 32: ......
Page 78: ...2 14 Connections ...
Page 98: ...3 20 Basic Source Measure Operation ...
Page 138: ...5 30 Source Measure Concepts ...
Page 156: ...6 18 Range Digits Speed and Filters ...
Page 168: ...7 12 Relative and Math ...
Page 176: ...8 8 Data Store ...
Page 202: ...9 26 Sweep Operation ...
Page 248: ...11 22 Limit Testing ...
Page 310: ...16 6 SCPI Signal Oriented Measurement Commands ...
Page 418: ...17 108 SCPI Command Reference ...
Page 450: ...18 32 Performance Verification ...
Page 477: ...A Specifications ...
Page 489: ...B StatusandErrorMessages ...
Page 498: ...B 10 Status and Error Messages ...
Page 499: ...C DataFlow ...
Page 503: ...D IEEE 488BusOverview ...
Page 518: ...D 16 IEEE 488 Bus Overview ...
Page 519: ...E IEEE 488andSCPI ConformanceInformation ...
Page 523: ...F MeasurementConsiderations ...
Page 539: ...G GPIB488 1Protocol ...
Page 557: ......