17-86
SCPI Command Reference
STATus subsystem
The STATus subsystem is used to control the status registers of the SourceMeter. The com-
mands in this subsystem are summarized in Table 17-7.
NOTE
These registers and the overall status structure are fully explained in Section 14.
Read event registers
[:EVENt]?
:STATus:MEASurement[:EVENt]?
Read Measurement Event Register
:STATus:QUEStionable[:EVENt]?
Read Questionable Event Register
:STATus:OPERation[:EVENt]?
Read Operation Event Register
Description
These query commands are used to read the contents of the status event reg-
isters. After sending one of these commands and addressing the SourceMe-
ter to talk, a value is sent to the computer. This value indicates which bits in
the appropriate register are set.
Program event enable registers
:ENABle <NDN> or <NRf>
:STATus:MEASurement:ENABle <NDN> or <NRf>
Program Measurement Event Enable Register
:STATus:QUEStionable:ENABle <NDN> or <NRf>
Program Questionable Event Enable Register
:STATus:OPERation:ENABle <NDN> or <NRf>
Program Operation Event Enable Register
Parameters
<NDN> =
#Bxx...x
Binary format (each x = 1 or 0)
#Hx
Hexadecimal format (x = 0 to FFFF)
#Qx
Octal format (x = 0 to 177777)
<NRf> =
0 to 65535
Decimal format
Query
:ENABle?
Read an enable register
Description
These commands are used to program the enable registers of the status
structure. The binary equivalent of the parameter value that is sent deter-
mines which bits in the register gets set. See Section 14 for details.
Summary of Contents for 6430
Page 26: ......
Page 32: ......
Page 78: ...2 14 Connections ...
Page 98: ...3 20 Basic Source Measure Operation ...
Page 138: ...5 30 Source Measure Concepts ...
Page 156: ...6 18 Range Digits Speed and Filters ...
Page 168: ...7 12 Relative and Math ...
Page 176: ...8 8 Data Store ...
Page 202: ...9 26 Sweep Operation ...
Page 248: ...11 22 Limit Testing ...
Page 310: ...16 6 SCPI Signal Oriented Measurement Commands ...
Page 418: ...17 108 SCPI Command Reference ...
Page 450: ...18 32 Performance Verification ...
Page 477: ...A Specifications ...
Page 489: ...B StatusandErrorMessages ...
Page 498: ...B 10 Status and Error Messages ...
Page 499: ...C DataFlow ...
Page 503: ...D IEEE 488BusOverview ...
Page 518: ...D 16 IEEE 488 Bus Overview ...
Page 519: ...E IEEE 488andSCPI ConformanceInformation ...
Page 523: ...F MeasurementConsiderations ...
Page 539: ...G GPIB488 1Protocol ...
Page 557: ......