SCPI Command Reference
17-41
2. If there are a lot of readings stored in the buffer, some statistic operations
may take too long and cause a bus time-out error. To avoid this, send the
:calc3:data? command and then wait for the MAV (message available) bit
in the Status Byte Register to set before addressing the SourceMeter to
talk. See Section 14.
3. See Appendix C for a detailed explanation on how data
fl
ows through the
various operation blocks of the SourceMeter. It clari
fi
es the types of read-
ings that are acquired by the various commands to read data.
:DISPlay subsystem
The display subsystem controls the display of the SourceMeter and is summarized in
Table 17-2.
Control display
:DIGits <n>
:DISPlay:DIGits <n>
Set display resolution
Parameters
<n> =
4
3.5 digit resolution
5
4.5 digit resolution
6
5.5 digit resolution
7
6.5 digit resolution
DEFault
5.5 digit resolution
MINimum
3.5 digit resolution
MAXimum
6.5 digit resolution
Query
:DIGits?
Query display resolution
:DIGits? DEFault
Query *RST default resolution
:DIGits? MINimum
Query lowest allowable display resolution
:DIGits? MAXimum
Query largest allowable display resolution
Description
This command is used to set the display resolution. Note that you can
instead use rational numbers. For example, to select 4.5 digit resolution, you
can send a parameter value of 4.5 (instead of 5). The SourceMeter rounds
the rational number to an integer.
Summary of Contents for 6430
Page 26: ......
Page 32: ......
Page 78: ...2 14 Connections ...
Page 98: ...3 20 Basic Source Measure Operation ...
Page 138: ...5 30 Source Measure Concepts ...
Page 156: ...6 18 Range Digits Speed and Filters ...
Page 168: ...7 12 Relative and Math ...
Page 176: ...8 8 Data Store ...
Page 202: ...9 26 Sweep Operation ...
Page 248: ...11 22 Limit Testing ...
Page 310: ...16 6 SCPI Signal Oriented Measurement Commands ...
Page 418: ...17 108 SCPI Command Reference ...
Page 450: ...18 32 Performance Verification ...
Page 477: ...A Specifications ...
Page 489: ...B StatusandErrorMessages ...
Page 498: ...B 10 Status and Error Messages ...
Page 499: ...C DataFlow ...
Page 503: ...D IEEE 488BusOverview ...
Page 518: ...D 16 IEEE 488 Bus Overview ...
Page 519: ...E IEEE 488andSCPI ConformanceInformation ...
Page 523: ...F MeasurementConsiderations ...
Page 539: ...G GPIB488 1Protocol ...
Page 557: ......