17-40
SCPI Command Reference
CALCulate3
Provides statistical data on buffer readings
Select statistic
:FORMat <name>
:CALCulate3:FORMat <name>
Specify CALC3 format
Parameters
<name> =
MEAN
Mean value of readings in buffer
SDEViation
Standard deviation of readings in buffer
MAXimum
Largest reading in buffer
MINimum
Lowest reading in buffer
PKPK
MAXimum - MINimum
Query
:FORMat?
Query programmed math format
Description
This command is used to select the desired statistic on readings stored in the
buffer. See Section 8 for details on these statistics.
Readings stored in the buffer can be
“
raw
”
measured readings, the results of
the CALC1 calculation, or CALC2 readings. The :TRACe:FEED command
in the :TRACe Subsystem is used to select the type of readings to store.
Acquire statistic
:DATA?
:CALCulate3:DATA?
Read CALC3 result
Description
This query command is used to perform the selected statistic operation and
read the result(s). The result(s) is always returned in ASCII format.
If the buffer is con
fi
gured to store
“
raw
”
measured readings (:TRACe:FEED
SENSe1) and multiple functions were measured, the selected statistic opera-
tion will be performed on all the measured readings. For example, if voltage
and current measurements were stored in the buffer, then the selected statis-
tic operation will be performed on both readings. Statistics for multiple
measurement functions are returned in the following order:
voltage statistic, current statistic, resistance statistic
Statistic operations are not performed on TIME and STATus data elements
that are stored in the buffer.
If the buffer is con
fi
gured to store the result of CALC1 or CALC2
(:TRACe:FEED CALC1 or CALC2), only one result will be returned by
this query command.
Notes:
1. If there is no data in the buffer, the NAN (not a number) value +9.91e37
will be returned.
Summary of Contents for 6430
Page 26: ......
Page 32: ......
Page 78: ...2 14 Connections ...
Page 98: ...3 20 Basic Source Measure Operation ...
Page 138: ...5 30 Source Measure Concepts ...
Page 156: ...6 18 Range Digits Speed and Filters ...
Page 168: ...7 12 Relative and Math ...
Page 176: ...8 8 Data Store ...
Page 202: ...9 26 Sweep Operation ...
Page 248: ...11 22 Limit Testing ...
Page 310: ...16 6 SCPI Signal Oriented Measurement Commands ...
Page 418: ...17 108 SCPI Command Reference ...
Page 450: ...18 32 Performance Verification ...
Page 477: ...A Specifications ...
Page 489: ...B StatusandErrorMessages ...
Page 498: ...B 10 Status and Error Messages ...
Page 499: ...C DataFlow ...
Page 503: ...D IEEE 488BusOverview ...
Page 518: ...D 16 IEEE 488 Bus Overview ...
Page 519: ...E IEEE 488andSCPI ConformanceInformation ...
Page 523: ...F MeasurementConsiderations ...
Page 539: ...G GPIB488 1Protocol ...
Page 557: ......