SCPI Command Reference
17-83
<NDN>
#Bx
Binary format:
3-bit: x = 000 to 111
4-bit: x = 0000 to 1111
#Hx
Hexadecimal format:
3-bit: x = 0 to 7
4-bit: x = 0 to F
#Qx
Octal format:
3-bit: x = 0 to 7
4-bit: x = 0 to 17
Query
:TTL?
Query digital output value
Description
This command is used to set the logic levels of the output lines of the Digital
I/O port. When set high, the speci
fi
ed output line will be at approximately
+5V. When set low, the output line will be at 0V.
Use the following table to determine the parameter value for the desired
decimal digital output pattern:
:ACTual?
:SOURce2:TTL[:LEVel]:ACTual?
Query bit pattern on digital output port.
Description
This query command is used to read the bit pattern that is actually on the
digital output port.
OUT 4
OUT 3
OUT 2
OUT 1
Decimal value*
L
L
L
L
L
L
L
L
H
H
H
H
H
H
H
H
L
L
L
L
H
H
H
H
L
L
L
L
H
H
H
H
L
L
H
H
L
L
H
H
L
L
H
H
L
L
H
H
L
H
L
H
L
H
L
H
L
H
L
H
L
H
L
H
0
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
L = Low (Gnd)
H = High (>+3V)
*0-7 in 3-bit mode.
Summary of Contents for 6430
Page 26: ......
Page 32: ......
Page 78: ...2 14 Connections ...
Page 98: ...3 20 Basic Source Measure Operation ...
Page 138: ...5 30 Source Measure Concepts ...
Page 156: ...6 18 Range Digits Speed and Filters ...
Page 168: ...7 12 Relative and Math ...
Page 176: ...8 8 Data Store ...
Page 202: ...9 26 Sweep Operation ...
Page 248: ...11 22 Limit Testing ...
Page 310: ...16 6 SCPI Signal Oriented Measurement Commands ...
Page 418: ...17 108 SCPI Command Reference ...
Page 450: ...18 32 Performance Verification ...
Page 477: ...A Specifications ...
Page 489: ...B StatusandErrorMessages ...
Page 498: ...B 10 Status and Error Messages ...
Page 499: ...C DataFlow ...
Page 503: ...D IEEE 488BusOverview ...
Page 518: ...D 16 IEEE 488 Bus Overview ...
Page 519: ...E IEEE 488andSCPI ConformanceInformation ...
Page 523: ...F MeasurementConsiderations ...
Page 539: ...G GPIB488 1Protocol ...
Page 557: ......