lxxx
Keysight E1412A Digital Multimeter User Guide
Keysight E1412A Multimeter Application Information
E1412A Multimeter Application Examples
End Loop
FETC?
Transfer measurements from the multimeter internal memory to the
output buffer and retrieve them with the computer.
Retrieve the
AVERage
math operation response from the multimeter.
CALC:AVER:AVER?
Retrieve the average measurement value.
CALC:AVER:MAX?
Retrieve the maximum measurement value
.
CALC:AVER:MIN?
Retrieve the minimum measurement value.
Check the multimeter for system errors.
SYST:ERR?
Retrieve the system error response from the multimeter.
SYNCHMAV Source Code File
This program has the multimeter take 10 measurements just like SYNCHOPC.
Readings are transferred to the output buffer by a
FETC?
command. The
Message Available bit (MAV) in the status byte (see Figure 2-5) is monitored to
detect when the measurements are complete and the Multimeter has readings in
the output buffer. Readings are retrieved by the computer when the MAV bit in
the status byte indicates the measurements are complete and readings are
available. The Multimeter then calculates the average, minimum and maximum
reading.
**** Set up the Multimeter ****
*RST
Reset the multimeter.
*CLS
Clear the multimeter's status registers.
CONF:VOLT 15
Configure for dc volts, expected input of 15V.
VOLT:DC:NPLC 10
Set number of power line cycles to 10.
TRIG:COUN 10
Multimeter will accept 10 triggers.
TRIG:DEL .01