153
2)
Send Instruction Format
3)
Return Instruction Format
The return instruction data is 0x00 or 0x01 or 0x02. When the return command data is 0x00, it means that
the time unit of the current step in the programmed mode is seconds; when the return command data is
0x01, it means that the time unit of the current step is the minute in the programmed mode; when the
return command data is 0x02, it means that the current mode is in the programmed mode. The time unit
of the step is time.
6.4.3.44
Address 44 (Delay Time)
1)
Command Description
This command is used to query the delay judgment time of the current step in the programmed mode.
2)
Send Instruction Format
3)
Return Instruction Format
The return instruction data range is from 0.1 to 999.9. When the time unit is second, the value is
0.1~999.9s; when the time unit is minute, the value is 0.1~999.9m; when the time unit is time, the value
is 0.1~999.9h.
Instrument
address
Function
Code
High Address
Low Address
High bytes
Low bytes
CRC Low
CRC High
1~32
0x03
0x00
0x2B
0x00
0x01
Calculated based on the first 6
bytes
Instrument
address
Function
Code
Number of
Data
Data Byte1
Data Byte2
CRC Low
CRC High
1~32
0x03
0x02
0x01
0x00
Calculated based on the first 5
bytes
0x01
0x02
Instrument
address
Function
Code
High Address
Low Address
High bytes
Low bytes
CRC Low
CRC High
1~32
0x03
0x00
0x2C
0x00
0x02
Calculated based on the first 6
bytes
Instrument
address
Function Code
Number of Data
Data Byte 1~4
CRC Low
CRC High
1~32
0x03
0x04
0.1~999.9, 4 bytes
stored in float point
type
Calculated based on the first 7 bytes