83
Shorthand for command: :FUNC:SD:VOLT:PROG <value>
Command description:This command is used to set the voltage value of the surge drop at the current step
in the programmable mode. <value> indicates the value of the surge drop
voltage, ranging from 0.0 to 300.0V. The command is invalid when the output
is turned on.
Command Syntax: :FUNCtion:SurgeDrop:VOLT:PROGram?
Shorthand for command: :FUNC:SD:VOLT:PROG?
Command description: This command is used to query the voltage value of the surge drop at the current
step in the programmable mode. Returns <value> with a value equal to 0.0~300.0V.
The command is invalid when the output is turned on.
Command Syntax: :FUNCtion:SurgeDrop:SITE:MANUal <value>
Shorthand for command: :FUNC:SD:SITE:MANU <value>
Command description: This command is used to set the position where the surge drop appears in manual
mode. <value> indicates the time at which the glitch trap appears, and its range is
0 to 20 ms.
Command Syntax: :FUNCtion:SurgeDrop:SITE:MANUal?
Shorthand for command: :FUNC:SD:SITE:MANU?
Command description: This command is used to query the location of the surge drop in manual mode.
Returns <value> with a value equal to 0~20.
Command Syntax: :FUNCtion:SurgeDrop:SITE:PROGram <value>
Shorthand for command: :FUNC:SD:SITE:PROG <value>
Command description: This command is used to set the position of the surge drop in the current step in
the programmable mode. <value> indicates the time at which the surge drop
appears, ranging from 0 to 20 ms. The command is invalid when the output is
turned on.
Command Syntax: :FUNCtion:SurgeDrop:SITE:PROGram?
Shorthand for command: :FUNC:SD:SITE:PROG?
Command description: This command is used to set the position of the surge drop in the current step in
the programmable mode. Returns <value> with a value equal to 0~20. The
command is invalid when the output is turned on.
Command Syntax: :FUNCtion:SurgeDrop:TIME:MANUal <value>
Shorthand for command: :FUNC:SD:TIME:MANU <value>
Command description: This command is used to set the time width of the glitch trap in manual mode.
<value> indicates the time width of the glitch trap, which ranges from 0 to 20 ms.
Command Syntax: :FUNCtion:SurgeDrop:TIME:MANUal?
Shorthand for command: :FUNC:SD:TIME:MANU?
Command description: This command is used to query the time width of the surge drop in manual mode.
Returns <value> with a value equal to 0~20.
Command Syntax: :FUNCtion:SurgeDrop:TIME:PROGram
<value>
Summary of Contents for TH7105
Page 14: ...9...