71
Command description: This command is used to set the number of cycles for the current step in the
programmable mode. <value> indicates the value, which ranges from 0 to 999.
6.3.5.5
:FUNC:VOLT
Command Syntax: :FUNCtion:VOLTage:MANUal
<value>
Shorthand for command: :FUNC:VOLT:MANU <value>
Command description: This command is used to set the set voltage value in manual mode. <value>
indicates the voltage value, and the range is 0.0~300.0V.
Command Syntax: :FUNCtion:VOLTage:MANUal?
Shorthand for command: :FUNC:VOLT:MANU?
Command description: This command is used to query the set voltage value in the current manual mode,
and returns <value>, which is equal to 0.0~300.0V.
Command Syntax: :FUNCtion:VOLTage:PROGram
<value>
Shorthand for command: :FUNC:VOLT:PROG <value>
Command description: This command is used to set the set voltage value of the current step in the
programmable mode. <value> indicates the voltage value, and the range is
0~300V. The command is invalid when the output is turned on.
Command Syntax: :FUNCtion:VOLTage:PROGram?
Shorthand for command: :FUNC:VOLT:PROG?
Command description: This command is used to query the set voltage value of the current step in the
current programmable mode, and returns <value>, which is equal to 0.0~300.0V.
The command is invalid when the output is turned on.
6.3.5.6
:FUNC:VOLT:HILMT
Command Syntax: :FUNCtion:VOLTage:HIghLiMiT:MANUal <value>
Shorthand for command: :FUNC:VOLT:HILMT:MANU <value>
Command description: This command is used to set the high voltage limit value in manual mode. <value>
indicates the high voltage limit value, and the range is 0.0~300.0V.
Command Syntax: :FUNCtion:VOLTage:HIghLiMiT:MANUal?
Shorthand for command: :FUNC:VOLT:HILMT:MANU?
Command description: This command is used to query the high voltage limit value in manual mode.
Command Syntax: :FUNCtion:VOLTage: HIghLiMiT:PROGram
<value>
Shorthand for command: :FUNC:VOLT:HILMT:PROG <value>
Command description: This command is used to set the high voltage limit value in the manual mode.
<value> indicates the high voltage limit value, and the range is 0.0~300.0V. The
command is invalid when the output is turned on.
Summary of Contents for TH7105
Page 14: ...9...