Command Syntax: :FUNCtion:DWELL?
Shorthand for command: :FUNC:DWELL?
Command description: This command is used to query the test time of the current step in the
programmable mode. Returns <value>, when the unit is second, its value is
0.1~999.9s; when the unit is minute, its value is 0.1~999.9m; when the unit is time,
its value is 0.1~999.9h. The command is invalid when the output is turned on.
6.3.5.23
:FUNC:RAMP
Command Syntax: :FUNCtion:RAMP:UP <value>
Shorthand for command: :FUNC:RAMP:UP <value>
Command description: This command is used to set the ramp up time of the current step in the
programmable mode. <value> indicates the ramp up time, which ranges from 0.0
to 999.9s. The command is invalid when the output is turned on.
Command Syntax: :FUNCtion:RAMP:UP?
Shorthand for command: :FUNC:RAMP:UP?
Command description: This command is used to query the ramp up time of the current step in the
programmable mode. Returns <value> with a value of 0.0~999.9s. The command
is invalid when the output is turned on.
Command Syntax: :FUNCtion:RAMP:DOWN <value>
Shorthand for command: :FUNC:RAMP:DOWN <value>
Command description: This command is used to set the ramp down time of the current step in the
programmable mode. <value> indicates the ramp down time, which ranges from
0.0 to 999.9s. The command is invalid when the output is turned on.
Command Syntax: :FUNCtion:RAMP:DOWN?
Shorthand for command: :FUNC:RAMP:DOWN?
Command description: This command is used to query the ramp down time of the current step in the
programmable mode. Returns <value> with a value of 0.0~999.9s. The command
is invalid when the output is turned on.
6.3.5.24
:FUNC:TIME
Command Syntax: :FUNCtion:TIME:SECond <value>
Shorthand for command: :FUNC: TIME:SEC <value>
Command description: This command is used to set the second value of the timer in manual mode.
<value> represents the timer seconds value, which ranges from 0 to 59.
Command Syntax: :FUNCtion:TIME:SECond?