74
0.000~8.400.
6.3.5.10
:FUNC:CURR:LOLMT
Command Syntax: :FUNCtion:CURRent:LoWLiMiT:MANUal <value>
Shorthand for command: :FUNC:CURR:LOLMT:MAUN <value>
Command description: This command is used to set the current low limit value in manual mode, and
<value> indicates the current low limit value. When the set voltage is not greater
than 150V, the setting range is 0.000~8.400A; when the set voltage is greater than
150V, the setting range is 0.000~4.200A. The command is invalid when the output
is turned on.
Command Syntax: :FUNCtion:CURRent:LoWLiMiT:MANUal?
Shorthand for command: :FUNC:CURR:LOLMT:MAUN?
Command description: This command is used to query the current high limit value in the current manual
mode, and returns <value>, which is equal to 0.000~8.400.
Command Syntax: :FUNCtion:CURRent:LOwLiMiT:PROGram <value>
Shorthand for command: :FUNC:CURR:LOLMT:PROG <value>
Command description: This command is used to set the current low limit value of the current step in the
programmable mode, and <value> indicates the current low limit value. When the
set voltage is not greater than 150V, the setting range is 0.000~8.400A; when the
set voltage is greater than 150V, the setting range is 0.000~4.200A. The command
is invalid when the output is turned on.
Command Syntax: :FUNCtion:CURRent:LOwLiMiT:PROGram?
Shorthand for command: :FUNC:CURR:LOLMT:PROG?
Command description: This command is used to query the current low limit of the current step in the
programmable mode, and returns <value>, which is equal to 0.000~8.400.
6.3.5.11
:FUNC:FREQ
Command Syntax: :FUNCtion:FREQuncy:MANUal <value>
Shorthand for command: :FUNC:FREQ:MANU <value>
Command description: This command is used to set the frequency value in manual mode, and <value>
represents the output frequency value. Its setting range is 45.0~500Hz. When the
frequency is less than 100 Hz, the resolution is 0.1 Hz; when the frequency is not
less than 100 Hz, the resolution is 1 Hz.
Command Syntax: :FUNCtion:FREQuncy:MANUal?
Shorthand for command: :FUNC:FREQ:MANU?
Command description:This command is used to query the output frequency value in manual mode, and