DISPlay Subsystem
DISPlay[:WINDow[1]|2]:ANALog:UPPer <numeric_value>
EPM-P Series Power Meters Programming Guide
5-13
Parameters
Example
Reset Condition
On reset, the upper scale limit is set to 20 dBm.
Query
DISPlay:[WINDow[1]|2]:ANALog:UPPer [MIN|MAX]
The query returns the current setting of the analog meter’s upper scale
limit, or the value associated with
MIN
or
MAX
. The format of the response
is
<NR3>
. The units in which the results are returned are determined by
the current setting of
UNIT:POWer
and
CALCulate:RELative:STATe
as
shown in Table 5-2.
Query Example
Item
Description/Default
Range of Values
numeric_value
A numeric value for the analog meter
upper scale limit:
•
DEF
: the default is 20 dBm
•
MIN
: -150 dBm
•
MAX
: 230 dBm
Units used are determined by the
current setting of
UNIT:POWer
and
CALCulate:RELative:STATe
as
shown in the previous table.
-150
to
230
dBm
DEF
MIN
MAX
DISP:WIND2:ANAL:UPP 50
This command sets the lower
window’s analog meter upper scale
limit to 50 dBm
DISP:WIND2:ANAL:UPP?
This command queries the upper
scale limit set on the analog meter
in the lower window