TRIGger Subsystem
TRIGger[:SEQuence]:LEVel:AUTO <boolean>
13-28
EPM-P Series Power Meters Programming Guide
TRIGger[:SEQuence]:LEVel:AUTO <boolean>
This command enables/disables automatic setting of the trigger level.
When this command is set to:
•
ON
, automatic setting of the trigger level is enabled.
•
OFF
, automatic setting of the trigger level is disabled.
•
ONCE
, automatic setting of the trigger level is enabled for one
trigger event only. The value is then set to
OFF
.
Note
The command is accepted for
TRIGger[:SEQuence[1]]
(channel A)
only, for both single and dual channel power meters.
Syntax
Example
Reset Condition
On reset the value is set to ON.
Query
TRIGger[:SEQuence]:LEVel:AUTO?
The query enters a
1
or
0
into the output buffer indicating the status of
TRIGger[:SEQuence]:LEVel:AUTO
.
•
1
is returned when it is
ON
.
•
0
is returned when it is
OFF
.
TRIG:SEQ:LEV:AUTO 0
This command disables the
automatic setting of the trigger
level for channel A.
:LEV
0|OFF
1|ON
Space
?
:AUTO
TRIG
:SEQ
ONCE