STATus Subsystem
Status Register Set Commands
10-4
EPM-P Series Power Meters Programming Guide
Status Register Set Commands
This section describes the five status register set commands. Each can be
used to examine all of the eleven status registers listed on the previous
page.
To apply a command to a specific register, prefix the command with the
name of the appropriate register. For example, to apply the
:ENABle
command to the
STATus:QUEStionable
register, use the following
command:
STATus:QUEStionable:ENABle
.
The Status Register Set commands detailed in this section are:
:CONDition?
This query returns a 16 bit decimal-weighted number representing the
bits set in the Condition Register of the SCPI Register Set you require to
control. The format of the return is
<NR1>
in the range of 0 to 32767
(2
15
-1). The contents of the Condition Register remain unchanged after it
is read.
Syntax
[:EVENt]?
This query returns a 16 bit decimal-weighted number representing the
bits set in the Event Register of the SCPI Register Set you require to
control. The format of the return is
<NR1>
in the range of 0 to 32767
(2
15
-1). This query clears all bits in the register to 0.
Syntax
Keyword
Parameter Form
Notes
Page
:CONDition?
[query only]
page 10-4
:ENABle
<NRf>|<non-decimal numeric>
page 10-5
[:EVENt?]
[query only]
page 10-4
:NTRansition
<NRf>|<non-decimal numeric>
page 10-6
:PTRansition
<NRf>|<non-decimal numeric>
page 10-7
:COND
?
:EVEN
?