Description of Functions
CT440
101
Configuration Functions
Parameters
iGPIBInterfaceID
GPIB interface ID of laser GPIB controller. Refer to your GPIB
controller firmware setup to set the correct ID number.
Possible values: 0 to 100
Type: Input
Data type: int32
iGPIBAdress
GPIB address of the laser.
Possible values: 1 to 30
Type: Input
Data type: int32
eLaserType
Laser type.
Possible values: see
Type: Input
Data type: rLaserSource
dMinWavelength
Laser minimum wavelength in nm.
Possible values: minimum of the wavelength range of the laser.
Type: Input
Data type: double
dMaxWavelength
Laser maximum wavelength in nm.
Possible values: maximum of the wavelength range of the laser.
Type: Input
Data type: double
Speed
Speed of laser in nm/s.
Possible values: from 10 to 100, limited to laser speed specifications.
Type: Input
Data type: int32
Return value
0: the laser setting operation succeeded.
-1: the laser setting operation failed.
Data type: int32
Example
CT440_SetLaser(uiHandle,LI_1,ENABLE,0,12,LS_T100S_HP,1480.0,160
0.0,100);