Description of Functions
108
CT440
Configuration Functions
CT440_SetBNC
Applicability
All models of CT440.
On CT440 with PDL option, PDL measurement is not guaranteed if
you use the BNC C interface.
Description
This function configures the CT440 BNC analog port labeled C
(located on the rear panel of the instrument) during the scan.
This function enables you to get the measured voltage converted in
dBm or mW, according to the entered parameters.
If this function is not called, the C detector can still be read; in this
case the returned values represent the voltage at the port (see
functions
CT440_ScanGetDetectorResampledArray
In the GUI, this function corresponds to the
Analog In
BNC
settings in
the
Referencing
window.
Declaration
int32_t CT440_SetBNC(uint64_t uiHandle,
rEnable eEnable,
double dAlpha,
double dBeta,
rUnit eUnit);
Parameters
uiHandle
Handle returned from
CT440_Init
Type: Input
Data type: unsigned int64
eEnable
Enables/Disables the laser output.
Possible values: see
Type: Input
Data type: rEnable
dAlpha
Conversion factor (slope) in mW/V or dBm
(P = dBeta + dAlpha x V)
Possible values: depend on the external power meter or detector
used.
Type: Input
Data type: double