Description of Functions
110
CT440
Configuration Functions
CT440_SetExternalSynchronization
Applicability
All models of CT440.
On CT440 with PDL option, you cannot use this function to perform a
direct PDL measurement. PDL measurements use resampled data,
not raw data.
Description
This function configures the CT440 external synchronization output
(
Trigger Out
BNC connector located on the rear panel of the
instrument).
The CT440 generates a pulse each time a measurement occurs.
For more details on this function, see
In the GUI, this function is always enabled.
Declaration
in32_t CT440_SetExternalSynchronization
(uint64_t uiHandle,
rEnable eEnable);
Parameters
uiHandle
Handle returned from
CT440_Init
Type: Input
Data type: unsigned int64
eEnable
Enables/Disables the laser output.
Possible values: see
Type: Input
Data type: rEnable
Return value
0: the operation succeeded.
-1: the operation failed.
Data type: int32
Example
CT440_SetExternalSynchronization(uiHandle,ENABLE);