7-28
Scanning
Model 2701 User’s Manual
Table 7-1
Scanning commands
Commands
Description
Default Ref
Scan commands
ROUTe:SCAN <clist>
Specify list of channels to be scanned.
a
ROUTe:SCAN?
Returns list of channels to be scanned.
ROUTe:SCAN:TSOurce <list>
Select trigger(s) to start scan; <list> = IMMediate or
HLIMit1, LLIMit1, HLIMit2, LLIMit2.
IMM
b
ROUTe:SCAN:NVOLatile <b>
Enable or disable auto scan; <b> = ON or OFF.
(Note 1)
ROUTe:SCAN:LSELect <name> Enable/disable scan; <name> = INTernal (on) or NONE
(off).
NONE
ROUTe:MONitor <clist>
Specify one channel to be monitored.
c
ROUTe:MONitor:POINts <NRf> Specify number of channels to scan; <NRf> = 2 to
450000.
(Note 3) d
ROUTe:MONitor:STATe <b>
Enable/disable monitor; <b> = ON or OFF.
OFF
ROUTe:MONitor:DATA?
Returns the most recent monitor reading.
ROUTe:CLOSe:ACONfigure <b> Enable or disable auto channel configuration.
Limits commands
CALCulate3:LIMit1:UPPer <n>
Set high limit 1 for monitor; <n> = -4294967295 to
+4294967295.
1.0
CALCulate3:LIMit1:LOWer <n> Set low limit 1 for monitor; <n> = -4294967295 to
+4294967295.
-1.0
CALCulate3:LIMit2:UPPer <n>
Set high limit 2 for monitor; <n> = -4294967295 to
+4294967295.
2.0
CALCulate3:LIMit2:LOWer <n> Set low limit 2 for monitor; <n> = -4294967295 to
+4294967295.
-2.0
CALCulate3:LIMit1:STATe <b>
Enable/disable limit 1 test; <b> = ON or OFF.
OFF
CALCulate3:LIMit2:STATe <b>
Enable/disable limit 2 test; <b> = ON or OFF.
OFF
Trigger commands
TRIGger:SOURce <name>
Select control source; <name> = IMMediate, TIMer,
MANual, BUS, or EXTernal.
IMM
TRIGger:TIMer <n>
Set timer interval in sec; <n> = 0.001 to 999999.999.
0.1
TRIGger:COUNt <NRf>
Set trigger count; <NRf> = 1 to 450000 or INFinity.
(Note 2) e
TRIGger:DELay <n>
Set delay in sec; <n> = 0 to 999999.999.
0
TRIGger:DELay:AUTO <b>
Enable/disable auto delay; <b> = ON or OFF.
ON
SAMPle:COUNt <NRf>
Set sample count; <NRf> = 1 to 450000.
1
e
SAMPle:COUNt?
Query sample count.
INITiate:CONTinuous <b>
Enable/disable continuous initiation; <b> = ON or OFF.
f
INITiate
Initiate one scan cycle.
f
READ?
Initiate one scan cycle and request sample readings.
f
2701-900-01.book Page 28 Wednesday, August 3, 2011 9:43 AM
Summary of Contents for 2701
Page 344: ...12 CommonCommands 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Page 366: ...15 SCPIReferenceTables 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Page 393: ...A Specifications 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Page 399: ...B Model7700ConnectionGuide 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Page 410: ...C StatusandErrorMessages 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Page 417: ...D SignalProcessing SequenceandDataFlow 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Page 430: ...E MeasurementConsiderations 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Page 450: ...G KE2700InstrumentDriver Examples 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...