12-6
Common Commands
Model 2701 User’s Manual
E *SAV <NRf> — save
Save present setup in memory
*RCL <NRf> — recall
Return to setup stored in memory
Parameters
0
= Memory
location
0
1
= Memory location 1
2
= Memory location 2
3
= Memory location 3
4
= Memory
location
4
Use the *SAV command to save the present instrument setup configuration in memory for
later recall. Any control affected by *RST can be saved by the *SAV command. The
*RCL command is used to restore the instrument to the saved setup configuration. Five
setup configurations can be saved and recalled. A saved setup is approximately 4k bytes in
size.
Model 2701 ships from the factory with SYSTem:PRESet defaults loaded into the
available setup memory. If a recall error occurs, the setup memory defaults to the
SYSTem:PRESet values.
NOTE
*OPC or *OPC? should be used with *SAV and *RCL, which are slow
responding commands. Details on *OPC and *OPC? are provided in this
section.
Programming example:
*SAV 2
' Save present setup in memory location 2.
*RST
' Return 2701 to RST defaults.
*RCL 2
' Return (recall) 2701 to setup stored in memo location 2.
F *RST — reset
Return Model 2701 to RST defaults
When the *RST command is sent, Model 2701 performs the following operations:
1.
Returns Model 2701 to the RST default conditions (see “Default” column of SCPI
tables).
2.
Cancels all pending commands.
3.
Cancels response to any previously received *OPC and *OPC? commands.
NOTE
*OPC or *OPC? should be used with *RST, which is a slow responding
command. Details on *OPC and *OPC? are provided in this section.
2701-900-01.book Page 6 Wednesday, August 3, 2011 9:43 AM
Summary of Contents for 2701
Page 344: ...12 CommonCommands 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Page 366: ...15 SCPIReferenceTables 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Page 393: ...A Specifications 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Page 399: ...B Model7700ConnectionGuide 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Page 410: ...C StatusandErrorMessages 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Page 417: ...D SignalProcessing SequenceandDataFlow 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Page 430: ...E MeasurementConsiderations 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Page 450: ...G KE2700InstrumentDriver Examples 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...